| Product Info |
 |
Custom Power Solution:
200W Dual Output Custom Power Inverter/Charger.
|
|
|
Features |
| AC feed through for normal operation |
| DC to AC converter when AC power is lost (105-120VAC square wave) |
| AC to DC battery charger (charges an external 42AH battery @ 12.6A) |
| Power factor correction(PFC) Meets EN60601-1-2 |
| FCC Class B EMI |
| No fans |
| Leakage current <100uA |
| |
| Input Characteristics |
Min |
Max |
Typical |
Units/Comments |
| Input Voltage Range |
105 |
120 |
|
VAC, 58-63 Hz |
| Efficiency Range: |
|
|
>89% Typ. FL |
|
| |
|
|
|
|
| Output Characteristics |
Min |
Max |
Typical |
Units/Comments |
| Output Power(Max.): |
|
200 |
|
Watts |
| Number of Outputs: |
|
2 |
|
|
| |
|
|
|
|
| Model |
Main Output Voltage (VDC) |
Aux 1 Nom. Vout (VDC) |
Aux 2 Nom. Vout (VDC) |
Main Output Current (Amps) |
Aux 1 Output Current (Amps) |
Aux 2 Output Current (Amps) |
| PS2401 |
15 |
|
|
12.6 |
|
|
|
| |
|
|
|
|
| Nominal Output VAC |
|
|
110 |
Vac, 105-120 |
| |
|
|
0 |
Battery 15V @ 12.6A |
| |
|
|
|
|
| Isolation Characteristics |
|
| Input to Output: |
4242 |
| |
|
|
|
|
| Environmental Characteristics |
|
| Operating Temperature: |
0 to +40 °C |
| Storage Temperature: |
-20 to +85 °C |
| Cooling: |
Convection |
| Humidity: |
Up to 95% non-condensing |
| |
|
|
|
|
| Mechanical Characteristics |
|
| Dimensions: |
6.0 x 2.8 x 13.9 inches
153.3 x 72 x 353 mm
|
| Case Material: |
Fully enclosed steel chassis |
| Indicators: |
Battery capacity / charging indicators (5 LEDs) |
| Connectors: |
Input & Output: SJT types wires |
| |
|
|
|
|
| Applicable Standards |
|
| EN60601-1-2 |
|
| EMI |
FCC Class B |
| UL |
|
| CSA |
|
| TUV |
|
| BART |
Pending |